TI1100 Precision AC DC Current Calibrator
TI1100 is a high-precision AC/DC current source designed for on-site calibration of ammeters and current measurement devices, making it ideal for industrial production lines, laboratories, and maintenance environments. It provides stable and accurate current outputs over a wide range, supporting both low and high current calibration tasks. The TI1100 achieves measurement accuracy up to Class 0.01 for DC current and Class 0.02 for AC current.
DC Current Output
| Range | Current Output | Resolution | Short-term Stability (%/min) | Accuracy (k=2) (%*output + μA) | Compliance Voltage (V) |
|---|---|---|---|---|---|
| 300 μA | ±10.0000 μA ~ 330.0000 μA | 100 pA | 0.02 | 0.012 + 0.03 | 11 |
| 3 mA | ±0.300000 mA ~ 3.300000 mA | 1 nA | 0.01 | 0.012 + 0.24 | 11 |
| 30 mA | ±3.00000 mA ~ 33.00000 mA | 10 nA | 0.01 | 0.012 + 2.4 | 9 |
| 300 mA | ±30.0000 mA ~ 330.0000 mA | 100 nA | 0.01 | 0.012 + 24 | 7 |
| 1 A | ±0.100000 A ~ 1.100000 A | 1 μA | 0.01 | 0.012 + 80 | 6 |
| 10 A | ±1.00000 A ~ 10.50000 A | 10 μA | 0.01 | 0.012 + 800 | 4 |
AC Current Output
| Range | Current Output | Frequency (Hz) | Resolution | Accuracy (k=2) (%*output + μA) | Compliance Voltage (V RMS) |
|---|---|---|---|---|---|
| 300 μA | 10.0000 μA ~ 330.0000 μA | 10 ~ 45 / 45 ~ 400 / 400 ~ 2000 | 100 pA | 0.05 + 0.2 / 0.03 + 0.1 / 0.04 + 0.2 | 7 |
| 3 mA | 0.300000 mA ~ 3.300000 mA | 10 ~ 45 / 45 ~ 400 / 400 ~ 2000 | 1 nA | 0.05 + 0.5 / 0.03 + 0.3 / 0.04 + 0.5 | 7 |
| 30 mA | 3.00000 mA ~ 33.00000 mA | 10 ~ 45 / 45 ~ 400 / 400 ~ 2000 | 10 nA | 0.05 + 5 / 0.03 + 3 / 0.04 + 5 | 6 |
| 300 mA | 30.0000 mA ~ 330.0000 mA | 10 ~ 45 / 45 ~ 400 / 400 ~ 2000 | 100 nA | 0.05 + 50 / 0.03 + 30 / 0.04 + 50 | 5 |
| 1 A | 0.100000 A ~ 1.100000 A | 10 ~ 45 / 45 ~ 400 / 400 ~ 2000 | 1 μA | 0.05 + 150 / 0.03 + 100 / 0.04 + 150 | 4 |
| 10 A | 1.00000 A ~ 10.50000 A | 45 ~ 400 | 10 μA | 0.03 + 1000 | 2.5 |
Note: RD = Reading Value, RG = Range Value.