TI2100 High Stability DC High Current Standard Source
TI2100 is a series of high-stability, precision DC high-current standard sources designed for industrial detection and calibration applications. With a modular architecture, the TI2100 supports combined outputs from multiple modules, allowing flexible adaptation to scenarios requiring different current specifications. It is ideal for calibrating DC shunts, current sensors, and other high-current measurement instruments in laboratories, production lines, and industrial environments.
The TI2100 delivers precise, stable current outputs with minimal ripple, fast settling times, and comprehensive protection features. Its high measurement accuracy and modular scalability make it suitable for both routine testing and high-precision calibration tasks.
TI2100 DC High Current Standard Source Specifications
| Parameter | Specification |
|---|---|
| Device Rating | Class 0.01 |
| Range | N × 1.5 kA |
| Output Range | (1% ~ 100%) × RG |
| Current Commutation | Supports external commutator (optional), supports bipolar output (optional) |
| Maximum Load Voltage | 10 V |
| Short-term Stability | 0.003% × FS |
| Measurement Uncertainty (k=2) | 60 ppm × RD + 40 ppm × RG |
| Ripple Coefficient | ≤ 0.1% |
| Settling Time | ≤ 1 s |
| Circuit Protection | Open circuit, overload, overtemperature protection |
| Note | RD = Reading Value, RG = Range Value |
General Specifications
| Power Supply | Three-phase five-wire, AC 380 V ± 38 V, 50 Hz ± 2 Hz |
| Weight | 84.5 kg (3 kA module) |
| Working Environment | 0°C ~ 50°C, 20% RH ~ 85% RH, non-condensing |
| Storage Environment | -20°C ~ 70°C, <85% RH, non-condensing |
| Communication Interface | LAN ×1, RS232 ×1, USB ×1 |
Note: RD = Reading Value, RG = Range Value, FS = Full Scale.